Skip to content
  • Basic research
  • Dental materials
  • Endodontics
  • Imaging
  • Implants
  • Infection control
  • Medically compromised and patients with special needs
  • Oral and maxillofacial surgery
  • Oral hygiene
  • Orthodontics
  • Other
  • Pediatric and preventive dentistry
  • Periodontics and oral medicine
  • Pharmacology
  • Prosthodontics
  • Restorative and aesthetic dentistry

Select Issue

  • Online First
  • 2024
    • Vol 28 – Issue 2
    • Vol 28 – Issue 1
  • 2023
    • VOL 27 – ISSUE 1
    • VOL 27 – ISSUE 2
    • VOL 27 – Issue 3
  • 2022
    • VOL 26 – ISSUE 1
    • VOL 26 – ISSUE 2
    • VOL 26 – ISSUE 3
  • 2021
    • VOL 25 – ISSUE 1
    • VOL 25 – ISSUE 2
    • VOL 25 – ISSUE 3
  • 2020
    • VOL 24 – ISSUE 1
    • VOL 24 – ISSUE 2
    • VOL 24 – ISSUE 3
  • 2019
    • VOL 23 – ISSUE 1
    • VOL 23 – ISSUE 2
    • VOL 23 – ISSUE 3
  • 2018
    • VOL 22 – ISSUE 1
    • VOL 22 – ISSUE 2
    • VOL 22 – ISSUE 3
  • 2017
    • VOL 21 – ISSUE 1
    • VOL 21 – ISSUE 2
    • VOL 21 – ISSUE 3
  • 2016
    • VOL 20 – ISSUE 1
    • VOL 20 – ISSUE 2
    • VOL 20 – ISSUE 3
  • 2015
    • VOL 19 – ISSUE 1
    • VOL 19 – ISSUE 2
    • VOL 19 – ISSUE 3
  • 2014
    • VOL 18 – ISSUE 1
    • VOL 18 – ISSUE 2
    • VOL 18 – ISSUE 3
  • Issue Archive
Skip to content
  • Balkan Stomatological Society
balkan.dentistry@hotmail.com
  • Facebook
Submit Manuscript
  • Facebook
Balkan Journal of Dental Medicine

Balkan Journal of Dental Medicine

Official Publication of the Balkan Stomatological Society

Submit Manuscript
  • Home
  • For Authors
    • Author Guidelines
    • Submit Your Manuscript
  • Articles and Issues
    • Online First
    • Issue Archive
      • 2026 YEAR
        • Vol 30 – ISSUE 1
      • 2025 YEAR
        • Vol 29 – ISSUE 1
        • Vol 29 – ISSUE 2
        • Vol 29 – ISSUE 3
      • 2024 YEAR
        • Vol 28 – Issue 3
        • Vol 28 – Issue 2
        • Vol 28 – Issue 1
      • 2023 YEAR
        • VOL 27 – ISSUE 1
        • VOL 27 – ISSUE 2
        • VOL 27 – ISSUE 3
      • 2022 YEAR
        • VOL 26 – ISSUE 1
        • VOL 26 – ISSUE 2
        • VOL 26 – ISSUE 3
      • 2021 YEAR
        • VOL 25 – ISSUE 1
        • VOL 25 – ISSUE 2
        • VOL 25 – ISSUE 3
      • 2020 YEAR
        • VOL 24 – ISSUE 1
        • VOL 24 – ISSUE 2
        • VOL 24 – ISSUE 3
      • 2019 YEAR
        • VOL 23 – ISSUE 1
        • VOL 23 – ISSUE 2
        • VOL 23 – ISSUE 3
      • 2018 YEAR
        • VOL 22 – ISSUE 1
        • VOL 22 – ISSUE 2
        • VOL 22 – ISSUE 3
      • 2017 YEAR
        • VOL 21 – ISSUE 1
        • VOL 21 – ISSUE 2
        • VOL 21 – ISSUE 3
      • 2016 YEAR
        • VOL 20 – ISSUE 1
        • VOL 20 – ISSUE 2
        • VOL 20 – ISSUE 3
      • 2015 YEAR
        • VOL 19 – ISSUE 1
        • VOL 19 – ISSUE 2
        • VOL 19 – ISSUE 3
      • 2014 YEAR
        • VOL 18 – ISSUE 1
        • VOL 18 – ISSUE 2
        • VOL 18 – ISSUE 3
      • Issue Archive
  • Categories
    • Oral and maxillofacial surgery
    • Periodontics and oral medicine
    • Pediatric and preventive dentistry
    • Restorative and aesthetic dentistry
    • Prosthodontics
    • Endodontics
    • Medically compromised and patients with special needs
    • Basic research
    • Orthodontics
    • Uncategorized
  • About Us
    • Aims and Scope
    • Editorial Board
    • Publication Policy
  • Contact Us
  • Menu Item

Accuracy of working length measurement by Raypex 6: Electronic apex locator versus actual measurements under stereomicroscope

Balkan Journal of Dental MedicineOctober 18, 2022January 24, 2023

Download Article

Peker Begum Unsal, Hepsenoglu Yelda Erdem, Ersahan Seyda, Eyuboglu Tan Firat

Istanbul Medipol University, Faculty of Dentistry, Department of Endodontics, Istanbul, Turkey

Summary

Background/Aim: This study aimed to evaluate the precision of the Raypex 6 apex locator in locating the apical constriction (AC) and major foramen (MF) during a root canal treatment compared with a microscopic evaluation. Another aim of the present study was to evaluate the accuracy of the Raypex 6 in the presence of different irrigating solutions (NaOCl, saline, EDTA, etidronic acid-HEBP). Material and Methods: One hundred and nine patients were included in this study and were randomly assigned into four groups according to the irrigation solution used; NaOCl, EDTA, HEBF and saline. Electronic MF (EMF) and AC (EAC) were measured by using Raypex 6. The teeth were extracted. The apical 3 mm of each canal was trimmed to expose the file tip. The samples were observed under a stereomicroscope, and the actual length of MF (ALMF) and AC (ALAC) were measured. The data were analyzed by using x2 test, and significance was set at P< 0.05. Results: The Raypex 6 was accurate 71.4% of the time to ±0.5mm and 93.3% of the time to ±1mm in determining the ALAC. While similar ALAC-EAC differences were observed in EDTA, NaOCl, and saline groups (p= 0.230), the highest differences were seen in the HEBP group (p= 1.000). The precision of Raypex 6 in determining the working length measurement depends on the type of irrigation. All solutions allowed reliable detection of AC. However, HEBF significantly increased the risk of overpreparation. Conclusions: Raypex 6 can be recommended for clinical use and its accuracy is not affected by the type of irrigant when locating MF.

Keywords: Raypex 6; Electronic Apex Locators; Working Length Measurement; Irrigation Solutions

Reference

Aydin, U., Karataslioglu, E., Aksoy, F., Yildirim, C. (2015) In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters. J Conserv Dent, 18(1): 66-69

Azabal, M., Garcia-Otero, D., de la Macorra, J.C. (2004) Accuracy of the Justy II Apex locator in determining working length in simulated horizontal and vertical fractures. Int Endod J, 37(3): 174-177

Brown, T.L., Lemay, H.E.Jr., Bursten, B.E., Murphy, C.J. (1997) Chemistry: The central science. J Chem Educ, 74: 378-379

Chukka, R.R., Bellam, M.D., Marukala, N.R., Dinapadu, S., Konda, N.K., Nagilla, J. (2020) Efficiency of an integrated apex locator in determining working length in various irrigating solutions: An in vivo study. J Pharm Bioallied Sci, 12(5): 410-414

Cianconi, L., Angotti, V., Felici, R., Conte, G., Mancini, M. (2010) Accuracy of Three Electronic Apex Locators Compared with Digital Radiography: An Ex Vivo Study. J Endod, 36(12): 2003-2007

Custer, L.E. (1918) Exact Methods of Locating the Apical Foramen. Journal of the National Dental Association, 5(8): 815-819

de Camargo, É.J., Zapata, R.O., Medeiros, P.L., Bramante, C.M., Bernardineli, N., Garcia, R.B., de Moraes, I.G., Duarte, M.A.H. (2009) Influence of Preflaring on the Accuracy of Length Determination With Four Electronic Apex Locators. J Endod, 35(9): 1300-1302

Dummer, P.M.H., Mcginn, J.H., Rees, D.G. (1984) The position and topography of the apical canal constriction and apical foramen. Int Endod J, 17(4): 192-198

Duran-Sindreu, F., Stöber, E., Mercadé, M., Vera, J., Garcia, M., Bueno, R., Roig, M. (2012) Comparison of In Vivo and In Vitro Readings When Testing the Accuracy of the Root ZX Apex Locator. J Endod, 38(2): 236-239

Ebrahim, A.K., Wadachi, R., Suda, H. (2006) Accuracy of three different electronic apex locators in detecting simulated horizontal and vertical root fractures. Aust Endod J, 32(2): 64-69

Elayouti, A., Kimionis, I., Chu, A., Löst, C. (2005) Determining the apical terminus of root-end resected teeth using three modern apex locators: A comparative ex vivo study. Int Endod J, 38(11): 827-833

Elayouti, A., Weiger, R., Lost, C. (2002) The Ability of Root ZX Apex Locator to Reduce the Frequency of Overestimated Radiographic Working length. J Endod, 28(2): 116-119

Erdemir, A., Eldeniz, A.U., Ari, H., Belli, S.H., Esener, T. (2007) The influence of irrigating solutions on the accuracy of the electronic apex locator facility in the Tri Auto ZX handpiece. Int Endod J, 40(5): 391-397

Goldberg, F., de Silvio, A.C., Manfré, S., Nastri, N. (2002) In vitro measurement accuracy of an electronic apex locator in teeth with simulated apical root resorption. J Endod, 28(6): 461-3

Gordon, M.P.J., Chandler, N.P. (2004) Electronic apex locators. Int Endod J, 37(7): 425-437

Hoer, D., Attin, T. (2004) The accuracy of electronic working length determination. Int Endod J, 37(2): 125-131

Jenkins, J.A., Walker, W., Schindler, W.G., Flores, C.M. (2001) An In Vitro Evaluation of the Accuracy of the Root ZX in the Presence of Various Irrigants. J Endod, 27(3): 209-211

Kim, E., Lee, S. (2004) Electronic apex locator. Dent Clin North Am, 48(1): 35-54

Krajczár, K., Marada, G., Gyulai, G., Tóth, V. (2008) Comparison of radiographic and electronical working length determination on palatal and mesio-buccal root canals of extracted upper molars. Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 106(2): e90-e93

Lee, S.J., Nam, K.C., Kim, Y.J., Kim, D.W. (2002) Clinical Accuracy of a New Apex Locator with an Automatic Compensation Circuit. J Endod, 28(10): 706-709

Lucena, C., López, J.M., Martín, J.A., Robles, V., González-Rodríguez, M.P. (2014) Accuracy of working length measurement: Electronic apex locator versus cone-beam computed tomography. Int Endod J, 47(3): 246-256

Mahmoud, O., Awad, A.M.H., Dandashi, A.H., Jasim, B.N., Tawfik, K.H.A., Al, S.S. (2021) Comparative Evaluation of Accuracy of Different Apex Locators: Propex IQ, Raypex 6, Root ZX, and Apex ID with CBCT and Periapical Radiograph-In Vitro Study. Int J Dent, 2021: 1-7

Malueg, L.A., Wilcox, L.R., Johnson, W. (1996) Examination of external apical root resorption with scanning electron microscopy. Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 82:89-93

Mancini, M., Felici, R., Conte, G., Costantini, M., Cianconi, L. (2011) Accuracy of Three Electronic Apex Locators in Anterior and Posterior Teeth: An Ex Vivo Study. J Endod, 37(5): 684-687

Martínez-Lozano, M.A., Forner-Navarro, L., Sánchez-Cortés, J.L., Llena-Puy, C. (2001) Methodological considerations in the determination of working length. Int Endod J, 34(5): 371-376

Milanović, I., Ivanović, V., Vujašković, M., Ignjatović, S., Miletić, V. (2015) Accuracy of three electronic apex locators in determining the apical foramen in multi-rooted teeth: Randomised clinical and laboratory study. Aust Endod J, 41(1): 35-43

Moscoso, S., Pineda, K., Basilio, J., Alvarado, C., Roig, M., Duran-Sindreu, F. (2014) Evaluation of Dentaport ZX and Raypex 6 electronic apex locators: An in vivo study. Med Oral Patol Oral Cir Bucal, 19: e202-e205

Ounsi, H.F., Naaman, A. (1999) In vitro evaluation of the reliability of the Root ZX electronic apex locator. Int Endod J, 32(2): 120-123

Pascon, E.A., Marrelli, M., Congi, O., Ciancio, R., Miceli, F., Versiani, M.A. (2009) Anin vivocomparison of working length determination of two frequency-based electronic apex locators. Int Endod J, 42(11): 1026-1031

Plotino, G., Grande, N.M., Brigante, L., Lesti, B., Somma, F. (2006) Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex. Int Endod J, 39(5): 408-414

Pratten, D.H., Mcdonald, N.J. (1996) Comparison of radiographic and electronic working lengths. J Endod, 22(4): 173-176

Ricucci, D., Langeland, K. (1998) Apical limit of root canal instrumentation and obturation: A histological study. Int Endod J, Part 2, 31(6): 394-409

Shabahang, S., Goon, W.W.Y., Gluskin, A.H. (1996) An in vivo evaluation of root ZX electronic apex locator. J Endod, 22(11): 616-618

Siu, C., Marshall, G.J., Baumgartner, C.J. (2009) An In Vivo Comparison of the Root ZX II, the Apex NRG XFR, and Mini Apex Locator by Using Rotary Nickel-Titanium Files. J Endod, 35(7): 962-965

Sjögren, U., Hägglund, B., Sundqvist, G., Wing, K. (1990) Factors affecting the long-term results of endodontic treatment. J Endod, 16(10): 498-504

Somma, F., Castagnola, R., Lajolo, C., Paternò, H.L., Marigo, L. (2012) In vivo accuracy of three electronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II. Int Endod J, 45(6): 552-556

Stein, T.J., Corcoran, J.F. (1992) Radiographic ‘working length’ revisited. Oral Surg Oral Med Oral Pathol, 74(6): 796-800

Stöber, E.K., Duran-Sindreu, F., Mercadé, M., Vera, J., Bueno, R., Roig, M. (2011) An Evaluation of Root ZX and iPex Apex Locators: An In Vivo Study. J Endod, 37(5): 608-610

Stöber, E.K., de Ribot, J., Mercadé, M., Vera, J., Bueno, R., Roig, M., Duran-Sindreu, F. (2011) Evaluation of the Raypex 5 and the Mini Apex Locator: An In Vivo Study. J Endod, 37(10): 1349-1352

Suzuki, K. (1942) Experimental study on iontophoresis. Jpn J Stomatol, 16: 411-429

Tselnik, M., Baumgartner, C.J., Marshall, G.J. (2005) An evaluation of root ZX and elements diagnostic apex locators. J Endod, 31(7): 507-9

Venturi, M., Breschi, L. (2005) A comparison between two electronic apex locators: An in vivo investigation. Int Endod J, 38(1): 36-45

Welk, A.R., Baumgartner, C.J., Marshall, G.J. (2003) An in vivo comparison of two frequency-based electronic apex locators. J Endod, 29(8): 497-500

Wrbas, K.T., Ziegler, A.A., Altenburger, M.J., Schirrmeister, J.F. (2007) In vivo comparison of working length determination with two electronic apex locators. Int Endod J, 40(2): 133-138

Wu, M.K., Wesselink, P.R., Walton, R.E. (2000) Apical terminus location of root canal treatment procedures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 89:99-103

Yolagiden, M., Ersahan, S., Suyun, G., Bilgec, E., Aydin, C. (2018) Comparison of Four Electronic Apex Locators in Detecting Working Length: An Ex Vivo Study. J Contemp Dent Pract, 19(12): 1427-1433

Balkan Journal of Dental Medicine, 2022, vol. 26, br. 1, str. 15-21

Post navigation

Previous: Three dimensional evaluation of marginal microleakage at the adhesive interface between different fissure sealants and enamel: Micro-CT analysis
Next: Root canal morphologies of mandibular premolars in Southeast Anatolian population: Evaluation of CBCT

Categories

  • Basic research
  • Dental materials
  • Endodontics
  • Imaging
  • Implants
  • Infection control
  • Medically compromised and patients with special needs
  • Oral and maxillofacial surgery
  • Oral hygiene
  • Orthodontics
  • Other
  • Pediatric and preventive dentistry
  • Periodontics and oral medicine
  • Pharmacology
  • Prosthodontics
  • Restorative and aesthetic dentistry
  • Facebook
  • Author Guidelines
  • Publication Policy
  • Editorial Board
  • Submit Manuscript
  • Contact us
All Rights Reserved | Theme: BlockWP by Candid Themes.
Go to Balkan Stomatological Society